Título: VALIDACIÓN DEL ESTUDIO DE REPETIBILIDAD Y REPRODUCIBILIDAD COMO PARTE DEL CONTROL METROLÓGICO EN EL LABORATORIO DE CONTROL DE CALIDAD DE UNA EMPRESA AGROINDUSTRIAL
VALIDATION STUDY OF REPEATABILITY AND REPRODUCIBILITY AS PART OF THE METROLOGICAL CONTROL IN THE LABORATORY OF QUALITY CONTROL OF AN AGROINDUSTRIAL COMPANY
ISSN: 2343-6115
Fecha: 2013
Páginas/Colación: pp. 38-62
En:/ Agroindustria, Sociedad y Ambiente Vol. 1 Nro. 1 Diciembre 2013
Información de existenciaInformación de existencia
Categoría Temática: Palabras: AGR02 AGR02
Palabras Claves del Autor: Palabras: CONTROL METROLÓGICO CONTROL METROLÓGICO, Palabras: REPETIBILIDAD REPETIBILIDAD, Palabras: REPRODUCIBILIDAD REPRODUCIBILIDAD
Documentos asociados
Oprima aquí para visualizar el documento pdf:Documento en Formato PDF (677027 bytes) Documento en Formato PDF (677027 bytes)

RESUMEN
La presente investigación tiene como principal objetivo validar el estudio de repetibilidad y reproducibilidad en los ensayos del laboratorio de control de calidad de una empresa agroindustrial como parte de un control metrológico. El desarrollo de la investigación se dividió en varias etapas: etapa de diagnóstico donde se aplicaron técnicas de recolección de datos como la observación directa, las encuestas y, la recolección y análisis de datos históricos aplicando estadística descriptiva, etapa de evaluación de los métodos para estudio de repetibilidad y reproducibilidad donde se seleccionó el más adecuado, etapa de esquematización del control metrológico que permitió lograr la estandarización del estudio de repetibilidad y reproducibilidad como parte del control metrológico y la etapa de aplicación del estudio que permitió determinar la confiabilidad de los ensayos realizados en el laboratorio. Por último se concluyó que el laboratorio de control de calidad no cumple con todos los requisitos establecidos por la Norma Técnica ISO 17025-2005, los analistas del laboratorio de control de calidad no poseen el conocimiento adecuado en materia metrológica, mientras que le jefa si, el estudio de repetibilidad y reproducibilidad puede aplicarse únicamente al análisis de humedad método rápido específicamente en las líneas PC1, PN5 y PL3, se seleccionó el método de promedios y rangos para el estudio R R, se logró estandarizar el control metrológico y finalmente los resultados de los ensayos realizados en el laboratorio no son confiables. La presente investigación aporta información vital para establecer un control metrológico que involucre estudios de repetibilidad y reproducibilidad.
Palabras Claves: Repetibilidad, reproducibilidad, control metrológico

ABSTRACT
This research has as main objective to validate the study of repeatability and reproducibility of laboratory testing for quality control of agroindustrial company as part of a metrological control. The development of the research was divided into several stages: diagnosis stage where data collection techniques were applied as direct observation , surveys , and the collection and analysis of historical data using descriptive statistics, step evaluation of methods to study repeatability and reproducibility which was selected the most appropriate, stage of outlining the metrological control that allowed to standardize the study of repeatability and reproducibility as part of metrological control and the implementation stage of the study that allowed us to determine the reliability of the tests performed laboratory . Finally it was concluded that the quality control laboratory does not meet all the requirements of the Technical Standard ISO 17025-2005 , analysts from laboratory quality control do not have adequate knowledge in metrology field , while the manager does , the study of repeatability and reproducibility can be applied only to the analysis of fast method moisture specifically on PC1, PN5 and PL3 lines , the method of averages and ranges for studying R R was selected, it was possible to standardize the metrological control and eventually the results of the tests performed in the laboratory are not reliable. This research provides vital information to establish metrological control involving studies of repeatability and reproducibility.
Keywords: repeatability, reproducibility, metrological control.

REFERENCIAS BIBLIOGRAFICAS
- Botero M, Arbeláez O, Mendoza J. (2007). Método ANOVA utilizado para realizar el estudio de repetibilidad y reproducibilidad dentro del control de calidad de un sistema de medición. Scientia et Technica. N° 37, p 533. Obtenido en Marzo 18 del 2012 en: http://www.utp.edu.co/php/revistas/ScientiaEtTechnica/docsFTP/155247533-537.pdf
- Briceño W. (2009). Desarrollo de pruebas de repetibilidad y reproducibilidad en análisis críticos del laboratorio de control de calidad de la empresa Levapan Venezolana S.A. Informe de pasantías. Universidad Centroccidental Lisandro Alvarado.
- Gutiérrez Pulido, H. De la Vara Salazar, R. (2009). Control Estadístico de la Calidad y Seis Sigma (2da ed.). Guzmán, E. (2011). Conceptos Básicos y Estadística Descriptiva. Obtenido el 20 de marzo del 2012 en: http://webdelprofesor.ula.ve/ingenieria/eliana/estad/unidad1EST.ppt
- C. Miller, J.N. Miller. (1988). Estadística para Química Analítica (Segunda Edición). Londres, Inglaterra: Addison-Wesley Iberoamerican, S.A.
- Llamosa L, Meza L, Botero M. (2007). Estudio de repetibilidad y reproducibilidad utilizando el método de promedios y rangos para el aseguramiento de la calidad de los resultados de calibración de acuerdo con la Norma Técnica NTC-ISO/IEC 17025. Scientia et Technica. N° 35, p - 455. Obtenido en Marzo 18 del 2012 en: revistas.utp.edu.co/
- Rodríguez, I. (2002). Control metrológico mediante la determinación de incertidumbre en las mediciones de los equipos de inspección, medición y ensayo de una empresa galvanizadora utilizando la Norma ISO 9001-2000. Universidad Nacional Abierta.


 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 